Intra-abdominal pressure measurements in lateral decubitus versus supine position.

De Keulenaer B L, Cheatham M L, De Waele J J, Kimball E J, Powell B, Davis W A, Jenkins I R

Контролируемое клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

Intra-abdominal pressure (IAP) has traditionally been measured in the supine position, however, measuring the pressure in lateral semi-recumbent position has not been studied.

DESIGN

A single centre prospective 1-day study.

PATIENTS: 10 patients admitted for more then 24 hours who were mechanically ventilated and had an indwelling urinary catheter.

METHODS

Inclusion criteria included were age > 18 years, sedated to a RASS score of -5 and mechanically ventilated. The pressures were measured via the bladder with the mid-axillary line as zero reference point. When patients were nursed in lateral decubitus, pressures were measured and compared immediately to the supine position.

RESULTS

10 patients were included with a total of 60 measurements.The male/female ratio was 9:1 with a mean APACHE Pi score of 11.5 [95% CI 4.8-22.4], SAPS 2 of 31.5 [95% CI 8.9-35.8] and SOFA score of 4.0 [95% CI 1.8-7.2]. Four patients were medical and 6 were surgical. The mean IAP at different time intervals (morning, afternoon and evening) in lateral and supine position were 10.9 +/- 2.0 (in mmHg) vs 6.6 +/- 3.2 (SD with p < 0.001); 11.0 +/- 4.0 vs 5.4 +/- 2.2 (p < 0.0005) and 11.6 +/- 3.8 vs 7.8 +/- 3.0 (p< 0.001). Overall, the LSP did not change significantly (p= 0.76), but the SP did (p=0.006) with the afternoon reading being significantly lower than the evening measurement. However, the trend in the difference (LSP minus SP) was not significant (p=0.43).

CONCLUSION

There was a significant statistical difference in the pressures measured in LSP versus SP.The LSP position should not be used to measure IAP.

Для цитирования: De Keulenaer B L, Cheatham M L, De Waele J J, Kimball E J, Powell B, Davis W A, Jenkins I R. Intra-abdominal pressure measurements in lateral decubitus versus supine position.. Acta clinica Belgica. 2009. DOI: 10.1179/acb.2009.037.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.