MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1464-410x.2009.08672.x

Медицинская библиотека · BJU international · 2009

Urinary catheter balloons should only be filled with water: testing the myth.

Huang James G, Ooi Jason, Lawrentschuk Nathan, Chan Steven T F, Travis Douglas, Wong Lih-Ming

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To test the hypothesis that urinary catheter balloons filled with sterile water, saline or glycine have equivalent rates of failure to deflate.

MATERIALS AND

METHODS

This was an in vitro equivalence study designed to test whether saline or glycine are neither substantially worse nor substantially better than water in terms of balloon-deflation failure rates. Glycine was chosen as the third arm, as it is readily available during endoscopic procedures and would be useful to use in such situations. We hypothesised that balloon-deflation failure rates using saline or glycine were no worse than water by 10%. We calculated the sample size for equivalence testing; 600 catheters were randomized by computer-generated random numbers to receive 10 mL of water, saline or glycine, and then immersed in a heated artificial urine solution for 6 weeks. The catheter balloons were then deflated, noting any failures to deflate and recording the deflation volumes.

RESULTS

There was no failure to deflate in all 600 catheters. The median deflation volume for water, saline and glycine was 9.0, 9.2 and 9.1 mL, respectively (P < 0.001 Kruskal-Wallis test). Post-hoc pair-wise comparisons showed that the deflation volume difference between water and saline was significant (P < 0.001), as was that between water and glycine (P < 0.001). The practical implication of this difference is not apparent from this study.

CONCLUSIONS

The use of saline or glycine in catheter balloons has an equivalent deflation failure rate to using water, which in this study was zero.

Для цитирования: Huang James G, Ooi Jason, Lawrentschuk Nathan, Chan Steven T F, Travis Douglas, Wong Lih-Ming. Urinary catheter balloons should only be filled with water: testing the myth.. BJU international. 2009. DOI: 10.1111/j.1464-410x.2009.08672.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.