VASA. Zeitschrift fur Gefasskrankheiten · 2011
Patient specific biomodel of the whole aorta - the importance of calcified plaque removal.
Клинический случайСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Håkansson A, Rantatalo M, Hansen T, Wanhainen A
Аннотация
BACKGROUND: The use of anatomical models produced by 3D printing technique (rapid prototyping, RP) is gaining increased acceptance as a complementary tool for planning complex surgical interventions. This paper describes a method for creating a patient specific replica of the whole aorta.
METHODS: Computed tomography angiography (CTA) DICOM-data was converted to a three-dimensional computer aided design-model (CAD) of the inner wall of the aorta representing the lumen where the calcified plaque contribution was removed in a multi-step editing-manoeuvre. The edited CAD-model was used for creating a physical plaster model of the true lumen in a 3D-printer. Elastic and transparent silicon was applied onto the plaster model, which was then removed leaving a silicon replica of the aorta.
RESULTS: The median (interquartile range) difference between diameters obtained from CTA- and RP plaster-model at 19 predefined locations was 0.5 mm (1 mm) which corresponds to a relative median difference of 4.6% (7.0%). The average wall thickness of the silicone model was 3.5 mm. The elasticity property and performance during intervention was good with an acceptable transparency.
CONCLUSIONS: The integration of RP-techniques with CAD based reconstruction of 3D-medical imaging data provides the needed tools for making a truly patient specific replica of the whole aorta with high accuracy. Plaque removal postprocessing is necessary to obtain a true inner wall configuration.
METHODS: Computed tomography angiography (CTA) DICOM-data was converted to a three-dimensional computer aided design-model (CAD) of the inner wall of the aorta representing the lumen where the calcified plaque contribution was removed in a multi-step editing-manoeuvre. The edited CAD-model was used for creating a physical plaster model of the true lumen in a 3D-printer. Elastic and transparent silicon was applied onto the plaster model, which was then removed leaving a silicon replica of the aorta.
RESULTS: The median (interquartile range) difference between diameters obtained from CTA- and RP plaster-model at 19 predefined locations was 0.5 mm (1 mm) which corresponds to a relative median difference of 4.6% (7.0%). The average wall thickness of the silicone model was 3.5 mm. The elasticity property and performance during intervention was good with an acceptable transparency.
CONCLUSIONS: The integration of RP-techniques with CAD based reconstruction of 3D-medical imaging data provides the needed tools for making a truly patient specific replica of the whole aorta with high accuracy. Plaque removal postprocessing is necessary to obtain a true inner wall configuration.
Предметные рубрики MeSH
AgedAortic DissectionAortaAortic AneurysmAortographyAtherosclerosisCalcinosisComputer SimulationComputer-Aided DesignFemaleHumansImage Processing, Computer-AssistedImaging, Three-DimensionalModels, AnatomicRadiographic Image EnhancementSensitivity and SpecificitySiliconesSoftwareTomography, Spiral Computed