The Laryngoscope · 2023
The Impact of Tracheal Stenosis on Distal Airway Pressure with Jet Ventilation.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Gottman Drew C, Corbisiero Michaele Francesco, Sus Ruba, Fink Daniel S
Аннотация
OBJECTIVE: This study investigates the effects of tracheal stenosis on distal airway pressure during low-frequency jet ventilation (LFJV) in tracheal stenosis resection procedures, focusing on variables like stenosis size, depth, scope type, and inlet pressure.
METHODS: A 3D-printed human airway model was employed, featuring inserted tracheal stenoses of varied sizes and depths. Distal airway pressure was measured with 16 pressure transducers, and data were processed via MATLAB. The study varied stenosis size, depth, scope type, and inlet pressure during five sequential jet bursts under LFJV.
RESULTS: Using a subglottiscope resulted in significantly reduced distal airway pressure compared to a laryngoscope. Interestingly, neither stenosis size nor depth significantly influenced distal airway pressure. However, increased distance between the scope and stenosis raised normalized pressure. A linear rise in normalized distal airway pressure was noted with increased inlet pressure, regardless of stenosis dimensions.
CONCLUSION: In this model, scope type and inlet pressure were noted to be significant determinants of distal airway pressure, while stenosis size and depth were not. The distance between the scope and the stenosis did influence distal pressures. These findings may have clinical implications for managing airway pressures in patients undergoing LFJV, potentially reducing the risk of ventilator-induced lung injury.
LEVEL OF EVIDENCE: NA (Basic Research) Laryngoscope, 134:2300-2305, 2024.
METHODS: A 3D-printed human airway model was employed, featuring inserted tracheal stenoses of varied sizes and depths. Distal airway pressure was measured with 16 pressure transducers, and data were processed via MATLAB. The study varied stenosis size, depth, scope type, and inlet pressure during five sequential jet bursts under LFJV.
RESULTS: Using a subglottiscope resulted in significantly reduced distal airway pressure compared to a laryngoscope. Interestingly, neither stenosis size nor depth significantly influenced distal airway pressure. However, increased distance between the scope and stenosis raised normalized pressure. A linear rise in normalized distal airway pressure was noted with increased inlet pressure, regardless of stenosis dimensions.
CONCLUSION: In this model, scope type and inlet pressure were noted to be significant determinants of distal airway pressure, while stenosis size and depth were not. The distance between the scope and the stenosis did influence distal pressures. These findings may have clinical implications for managing airway pressures in patients undergoing LFJV, potentially reducing the risk of ventilator-induced lung injury.
LEVEL OF EVIDENCE: NA (Basic Research) Laryngoscope, 134:2300-2305, 2024.