DOI 10.1067/mje.2003.37

Медицинская библиотека · Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography · 2003

Real-time three-dimensional color doppler evaluation of the flow convergence zone for quantification of mitral regurgitation: Validation experimental animal study and initial clinical experience.

Sitges Marta, Jones Michael, Shiota Takahiro, Qin Jian Xin, Tsujino Hiroyuki, Bauer Fabrice, Kim Yong Jin, Agler Deborah A, Cardon Lisa A, Zetts Arthur D, Panza Julio A, Thomas James D, New Collective Author

Клиническое исследованиеСравнительное исследованиеОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПоддержка правительства СШАВалидационное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Pitfalls of the flow convergence (FC) method, including 2-dimensional imaging of the 3-dimensional (3D) geometry of the FC surface, can lead to erroneous quantification of mitral regurgitation (MR). This limitation may be mitigated by the use of real-time 3D color Doppler echocardiography (CE). Our objective was to validate a real-time 3D navigation method for MR quantification.

METHODS

In 12 sheep with surgically induced chronic MR, 37 different hemodynamic conditions were studied with real-time 3DCE. Using real-time 3D navigation, the radius of the largest hemispherical FC zone was located and measured. MR volume was quantified according to the FC method after observing the shape of FC in 3D space. Aortic and mitral electromagnetic flow probes and meters were balanced against each other to determine reference MR volume. As an initial clinical application study, 22 patients with chronic MR were also studied with this real-time 3DCE-FC method. Left ventricular (LV) outflow tract automated cardiac flow measurement (Toshiba Corp, Tokyo, Japan) and real-time 3D LV stroke volume were used to quantify the reference MR volume (MR volume = 3DLV stroke volume - automated cardiac flow measurement).

RESULTS

In the sheep model, a good correlation and agreement was seen between MR volume by real-time 3DCE and electromagnetic (y = 0.77x + 1.48, r = 0.87, P <.001, delta = -0.91 +/- 2.65 mL). In patients, real-time 3DCE-derived MR volume also showed a good correlation and agreement with the reference method (y = 0.89x - 0.38, r = 0.93, P <.001, delta = -4.8 +/- 7.6 mL).

CONCLUSIONS

real-time 3DCE can capture the entire FC image, permitting geometrical recognition of the FC zone geometry and reliable MR quantification.

Для цитирования: Sitges Marta, Jones Michael, Shiota Takahiro, Qin Jian Xin, Tsujino Hiroyuki, Bauer Fabrice, Kim Yong Jin, Agler Deborah A, Cardon Lisa A, Zetts Arthur D, Panza Julio A, Thomas James D, New Collective Author. Real-time three-dimensional color doppler evaluation of the flow convergence zone for quantification of mitral regurgitation: Validation experimental animal study and initial clinical experience.. Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography. 2003. DOI: 10.1067/mje.2003.37.