MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jvs.2017.07.133

Медицинская библиотека · Journal of vascular surgery · 2017

Ten-year outcome analysis of the Italian Excluder Registry with the Gore Excluder endograft for infrarenal abdominal aortic aneurysms.

Pratesi Giovanni, Piffaretti Gabriele, Verzini Fabio, De Blasis Giovanni, Castelli Patrizio, Pratesi Carlo

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVE

The objective of this study was to report the 10-year outcomes of the multicenter Italian Excluder Registry for elective endovascular aneurysm repair (EVAR).

METHODS

A retrospective analysis of all patients who underwent elective EVAR using the Gore Excluder (W. L. Gore & Associates, Flagstaff, Ariz) endograft (EG) at four Italian centers between 1998 and 2006 was carried out. Follow-up consisted of duplex ultrasound scan and computed tomography performed at 1 month, 6 months, and 12 months and yearly thereafter. Long-term outcomes were evaluated according to a life-table analysis.

RESULTS

The Excluder EG was used in 461 patients (425 men; mean age, 72.9 ± 9.2 years) with abdominal aortic aneurysm (mean diameter, 52.2 ± 11.9 mm) who underwent elective EVAR. The 10-year cumulative survival was 62.5% ± 3.5% (95% confidence interval [CI], 55.5%-69.1%). During the follow-up, 14 (3.03%) patients were lost to follow-up. Estimated freedom from EG-related complication was 90.5% ± 1.5% (95% CI, 87.4%-92.9%) at 5 years, 89.2% ± 1.6% (95% CI, 85.6%-91.9%) at 7 years, and 88.4% ± 1.8% (95% CI, 84.4%-91.5%) at 10 years. Freedom from reintervention at 5, 7, and 10 years was 87.7% ± 1.8% (95% CI, 83.7%-90.8%), 82.4% ± 2.4% (95% CI, 77.2%-86.6%), and 80.6% ± 2.6% (95% CI, 75.5%-84.9%), respectively. On multivariable analysis, we did not find independent predictors of a higher rate of reintervention. There were 127 (27.5%) endoleaks detected during the follow-up. Only one type I endoleak was identified beyond 5-year follow-up. Cumulative freedom from endoleak rate was significantly different between the two different follow-up intervals (0-5 years, 2.6%; 6-10 years, 0.8%; P < .001). Estimated freedom from aorta-related mortality was 97.2% ± 0.8% (95% CI, 49.7%-99.9%) at 10 years.

CONCLUSIONS

The Italian Excluder Registry outcomes confirmed sustained EVAR effectiveness at 10-year follow-up using the Gore Excluder EG. Freedom from EG-related complications, endoleaks, and reinterventions was satisfactory and remained acceptable beyond 5 years of follow-up, supporting a reasonable long-term durability of this device.

Для цитирования: Pratesi Giovanni, Piffaretti Gabriele, Verzini Fabio, De Blasis Giovanni, Castelli Patrizio, Pratesi Carlo. Ten-year outcome analysis of the Italian Excluder Registry with the Gore Excluder endograft for infrarenal abdominal aortic aneurysms.. Journal of vascular surgery. 2017. DOI: 10.1016/j.jvs.2017.07.133.