MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1161/jaha.117.006265

Медицинская библиотека · Journal of the American Heart Association · 2017

Low Prevalence of Inappropriate Shocks in Patients With Inherited Arrhythmia Syndromes With the Subcutaneous Implantable Defibrillator Single Center Experience and Long-Term Follow-Up.

Rudic Boris, Tülümen Erol, Berlin Veronika, Röger Susanne, Stach Ksenija, Liebe Volker, El-Battrawy Ibrahim, Dösch Christina, Papavassiliu Theano, Akin Ibrahim, Borggrefe Martin, Kuschyk Jürgen

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Up to 40% of patients with transvenous implantable cardioverter-defibrillator (ICD) experience lead-associated complications and may suffer from high complication rates when lead extraction is indicated. Subcutaneous ICD may represent a feasible alternative; however, the efficacy of the subcutaneous ICD in the detection and treatment of ventricular arrhythmias in patients with hereditary arrhythmia syndromes has not been fully evaluated.

METHODS AND

RESULTS

Patients with primary hereditary arrhythmia syndromes who fulfilled indication for defibrillator placement were eligible for enrollment. Between 2010 and 2016, 62 consecutive patients with primary hereditary arrhythmia syndromes, without indication for antibradycardia therapy, were enrolled in the study. Mean follow-up was 31.0±14.2 months. The study cohort comprised of 24 patients with Brugada syndrome, 17 with idiopathic ventricular fibrillation, 6 with long-QT syndrome, 1 with short-QT syndrome, 3 with catecholaminergic polymorphic ventricular tachycardia, 8 with hypertrophic cardiomyopathy, and 3 with arrhythmogenic right ventricular cardiomyopathy. Thirty-nine patients were implanted for secondary prevention. Twenty-two patients had a previous transvenous ICD implanted, but required revision because of infection or lead defects. A total of 20 spontaneous ventricular tachyarrhythmias requiring shock intervention occurred in 10 patients during follow-up. All episodes were terminated within the first ICD shock delivery with 80 J. Two patients had inappropriate therapies caused by oversensing following an uneventful implantation. No pocket-site infections and no premature revisions have occurred during follow-up.

CONCLUSIONS

Our study supports the use of the subcutaneous ICD for both secondary and primary prevention of sudden cardiac death as a reliable alternative to the conventional transvenous ICD.

Для цитирования: Rudic Boris, Tülümen Erol, Berlin Veronika, Röger Susanne, Stach Ksenija, Liebe Volker, El-Battrawy Ibrahim, Dösch Christina, Papavassiliu Theano, Akin Ibrahim, Borggrefe Martin, Kuschyk Jürgen. Low Prevalence of Inappropriate Shocks in Patients With Inherited Arrhythmia Syndromes With the Subcutaneous Implantable Defibrillator Single Center Experience and Long-Term Follow-Up.. Journal of the American Heart Association. 2017. DOI: 10.1161/jaha.117.006265.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.