MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/1457496916683093

Медицинская библиотека · Scandinavian journal of surgery : SJS : official organ for the Finnish Surgical Society and the Scandinavian Surgical Society · 2017

Systematic Screening of Adverse Reactions to Metal Debris after Recap-M2A-Magnum Metal-on-Metal Total Hip Arthroplasty.

Mäntymäki H, Junnila M, Lankinen P, Seppänen M, Vahlberg T, Mäkelä K T

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND AIMS: An adverse reaction to metal debris is a known complication after large diameter head metal-on-metal total hip arthroplasty. However, the failure rate varies depending on the implant design. Therefore, we investigated the prevalence of adverse reaction to metal debris, as well as the symptoms and risk factors after undergoing a ReCap-M2a-Magnum large diameter head metal-on-metal total hip arthroplasty.

MATERIALS AND

METHODS

Between 2005 and 2012, 1188 patients (1329 hips) underwent ReCap-M2a-Magnum total hip arthroplasty at our institution. Systematic screening for adverse reaction to metal debris was arranged using the Oxford Hip Score questionnaire, hip and pelvic radiographs, and assessments of the serum chromium and cobalt ion levels. Clinical evaluation and magnetic resonance imaging were performed for the symptomatic patients, as well as those with either chromium or cobalt ion levels ⩾5 µg/L. The prevalence of adverse reaction to metal debris after ReCap-M2a-Magnum total hip arthroplasty was assessed, and the risk factors for adverse reaction to metal debris were evaluated using logistic regression. The mean follow-up time was 5.2 (0.003-9.1) years. This study was an extension of a previous study conducted at our institution with 80 patients.

RESULTS

In total, 33 patients (33 hips, 2.5% of all hips) required a revision operation due to adverse reaction to metal debris. Moreover, 157 hips exhibited definitive adverse reaction to metal debris, but a revision operation was not performed (157 of 1329 hips, 11.8% of all hips). Overall, 190 out of 1329 (14.3%) hips had definitive adverse reaction to metal debris. Pain, subluxation sensation, clicking, swelling, a small head size, and a fair/poor Oxford Hip Score were associated with definitive adverse reaction to metal debris.

CONCLUSION

We found a high prevalence of adverse reaction to metal debris in the ReCap-M2a-Magnum total hip arthroplasty patients in this study; however, most of the patients did not require revision operations.

Для цитирования: Mäntymäki H, Junnila M, Lankinen P, Seppänen M, Vahlberg T, Mäkelä K T. Systematic Screening of Adverse Reactions to Metal Debris after Recap-M2A-Magnum Metal-on-Metal Total Hip Arthroplasty.. Scandinavian journal of surgery : SJS : official organ for the Finnish Surgical Society and the Scandinavian Surgical Society. 2017. DOI: 10.1177/1457496916683093.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.