Медицинская библиотека · Clinical medicine insights. Case reports · 2019
Unfortunate Hand Positioning Yielding Inappropriate Defibrillator Shock: Another Paradigm of a Guilty Association Between Patient Awareness and Occupational Hazard.
Клинический случайСтатьяПеревод пока не подготовлен
Аннотация
A 64-year-old patient had a cardiac resynchronization therapy defibrillator implanted 4 years ago for dilated cardiomyopathy. He recently had an inappropriate shock as a result of an electrical leak from a refrigerator causing electromagnetic interference. An analysis of the patient case and a discussion of the features of this specific paradigm of inappropriate therapy are presented.
Для цитирования: Kossaify Antoine. Unfortunate Hand Positioning Yielding Inappropriate Defibrillator Shock: Another Paradigm of a Guilty Association Between Patient Awareness and Occupational Hazard.. Clinical medicine insights. Case reports. 2019. DOI: 10.1177/1179547619861096.