DOI 10.1111/nmo.14505

Медицинская библиотека · Neurogastroenterology and motility · 2022

Inter- and intra-rater agreement of interpretation of functional lumen imaging probe in healthy subjects.

Yadlapati Rena, Gyawali C Prakash, Carlson Dustin A, Pandolfino John E, Fass Ronnie, Khan Abraham, Lin Haiying, Richter Joel E, Vela Marcelo F, Vaezi Michael, Clarke John O

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The functional lumen imaging probe (FLIP) evaluates esophagogastric junction (EGJ) opening and esophageal contractility. Both post hoc and real-time analyses are possible, but reproducibility and reliability of analysis remain undefined. This study assesses inter- and intra-rater agreement of normative FLIP measurements among novice and experienced users.

METHODS

Eight motility experts from different institutions independently evaluated de-identified video recordings from 27 asymptomatic healthy subjects using FLIP. Interpretation methods simulating a post-procedure and a live procedure setting were tested. Novice FLIP users (n = 3) received training prior to post-procedure interpretation. Experienced FLIP users (n = 5) interpreted using both methods. Users recorded maximum EGJ and distal esophageal body diameter, distensive pressure, and EGJ distensibility index (EGJ-DI), at balloon fill volumes of 50-, 60-, and 70 ml, as well as repetitive antegrade contractions (RACs). Inter- and intra-rater agreements of diameters, distensive pressure and EGJ-DI were assessed by intra-class correlation coefficient (ICC) and Pearson's correlation coefficient (PCC). Percentage agreement evaluated inter- and intra-rater reliability for RACs.

KEY

RESULTS

Novice and experienced users acquired normative FLIP metrics. Good-to-excellent inter- and intra-rater reliability were achieved for all variables at 60 ml balloon fill volumes. Median parameters at 60 ml balloon fill volume were as follows: EGJ-DI 5.5 mm2 /mmHg, maximum EGJ diameter 18.6 mm, distensive pressure at maximum EGJ diameter 48.1 mmHg, and distal esophageal body diameter 19.5 mm.

CONCLUSIONS AND INFERENCES: Normative FLIP parameters can be reliably extracted from FLIP videos using both real-time and post hoc analyses, with high reliability between experienced and novice users.

Для цитирования: Yadlapati Rena, Gyawali C Prakash, Carlson Dustin A, Pandolfino John E, Fass Ronnie, Khan Abraham, Lin Haiying, Richter Joel E, Vela Marcelo F, Vaezi Michael, Clarke John O. Inter- and intra-rater agreement of interpretation of functional lumen imaging probe in healthy subjects.. Neurogastroenterology and motility. 2022. DOI: 10.1111/nmo.14505.