MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/bf03018485

Медицинская библиотека · Canadian journal of anaesthesia = Journal canadien d'anesthesie · 2004

A new guidance device facilitates percutaneous puncture of the foramen ovale in human cadavers.

Hinteregger Martin, Zschiegner Fritz, Lirk Philipp, Ladner Eugen, Goeschl Albert, Gaber Othmar, Moser Patrizia, Lorenz Ingo, Kolbitsch Christian

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Trigeminal neuralgia is the most common neurological cause for facial pain. Contemporary interventional treatment relies on surgical microvascular decompression or, alternatively, percutaneous interventions targeting the semilunar ganglion via the foramen ovale. For the latter approach, only free-hand punctures using fluoroscopy devices have been reported. Therefore, the present study aimed to evaluate a new fluoroscopy-based guidance device for transforaminal puncture.

METHODS

Two experienced examiners punctured the foramen ovale bilaterally free-hand, and using a guidance device in human cadavers (n = 9). The number of attempts for puncture was recorded. A new attempt was counted each time the needle had to be retracted for redirection.

RESULTS

As compared to the free-hand puncture of the foramen ovale (4.44 +/- 2.79), the new guidance device significantly reduced the number of trials needed (1.37 +/- 0.69).

CONCLUSION

The employment of a guidance device facilitated percutaneous transforaminal puncture and resulted in a significantly decreased number of puncture attempts as compared to free-hand techniques in human cadavers.

Для цитирования: Hinteregger Martin, Zschiegner Fritz, Lirk Philipp, Ladner Eugen, Goeschl Albert, Gaber Othmar, Moser Patrizia, Lorenz Ingo, Kolbitsch Christian. A new guidance device facilitates percutaneous puncture of the foramen ovale in human cadavers.. Canadian journal of anaesthesia = Journal canadien d'anesthesie. 2004. DOI: 10.1007/bf03018485.