MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1117/1.jbo.25.10.106002

Медицинская библиотека · Journal of biomedical optics · 2020

USB capsule endoscope for retrograde imaging of the esophagus.

Martincek Ivan, Banovcin Peter, Goraus Matej, Duricek Martin

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

SIGNIFICANCE: Endoscopes represent electro-optical devices that are used to visualize internal body cavities. The specialized endoscopic procedure of the upper gastrointestinal tract from the esophagus down to the duodenum is called an esophagogastroduodenoscopy.

AIM: We bring our newly developed capsule endoscopy device as a promising alternative diagnostic method for visualization of the upper gastrointestinal tract.

APPROACH: Capsule endoscopy has become an attractive method that uses a tiny wireless camera to take pictures of the digestive tract. Existing esophageal capsule endoscopy does not allow a retrograde view of the esophagus while retrograde scanning can provide information on the esophageal pathology.

RESULTS

In comparison to the existing esophageal capsule endoscopy, our system is much simpler and cheaper due to the need for fewer electronic devices. Moreover, its use is not limited by the capacity of the batteries used by existing capsule endoscopes. The new esophageal endoscopic system was created by combining the universal serial bus (USB) endoscope module with the thin power wires that are routed through the USB port to the computer.

CONCLUSIONS

The endoscope was tested on a volunteer without any side effects such as nausea, belching, and general discomfort. The examination of the patient is performed in a sitting position and the patient discomfort during the examination is minimal so it can be performed without anesthesia.

Для цитирования: Martincek Ivan, Banovcin Peter, Goraus Matej, Duricek Martin. USB capsule endoscope for retrograde imaging of the esophagus.. Journal of biomedical optics. 2020. DOI: 10.1117/1.jbo.25.10.106002.