DOI 10.1016/j.vgie.2025.07.011
Clinical utility of a novel double-band endoscopic band ligation device for consecutive ligation in the management of diverticular bleeding.
Аннотация
BACKGROUND AND AIMS: Endoscopic band ligation is effective for diverticular bleeding; however, conventional devices have drawbacks such as limited visibility, weak suction, and single-use design. Endoligar (Asahi Intecc Co Ltd, Aichi, Japan) addresses these limitations with clear visibility, strong suction, and the ability to perform 2 ligations without removing the endoscope. This report evaluates its clinical utility.
METHODS
An 82-year-old man with a history of diverticular bleeding was admitted for hematochezia lasting 3 days. Noncontrast computed tomography and emergency colonoscopy suggested a bleeding diverticulum in the ascending colon. A diverticulum at the hepatic flexure showed signs of a recent hemorrhage, but no active bleeding; a clip was placed for marking. On day 8, a second colonoscopy showed oozing from the previously marked diverticulum, and endoscopic band ligation was performed using the novel device. The first band dislodged, but a second ligation was completed without scope removal.
RESULTS
The procedure lasted 27 minutes, with minimal bleeding and no adverse events. The patient was discharged on day 13.
CONCLUSIONS
This is the first clinical case report on this device, to our knowledge. The device enables a second ligation without scope removal, reducing the risk of losing the lesion, and is potentially useful in challenging endoscopic band ligation cases. Further studies are required.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.