Comparison of three cough-augmentation techniques in neuromuscular patients: mechanical insufflation combined with manually assisted cough, insufflation-exsufflation alone and insufflation-exsufflation combined with manually assisted cough.

Lacombe Matthieu, Del Amo Castrillo Lorena, Boré Aurélien, Chapeau David, Horvat Eric, Vaugier Isabelle, Lejaille Michelle, Orlikowski David, Prigent Hélène, Lofaso Frédéric

Сравнительное исследованиеСтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Mechanical insufflation-exsufflation (MI-E), more commonly known as 'cough assist therapy', is a method which produces inspiratory and expiratory assistance to improve cough performances. However, other alternatives or combinations are possible.

OBJECTIVE

The objective was to compare the effects of mechanical insufflation combined with manually assisted coughing (MAC), insufflation-exsufflation alone and insufflation-exsufflation combined with MAC in neuromuscular patients requiring cough assistance.

METHODS

Eighteen neuromuscular patients with severe respiratory muscle dysfunction and peak cough flow (PCF) lower than 3 liters/s or maximal expiratory pressure (MEP) lower than +45 cm H2O were studied. Patients were studied under three cough-assisted conditions, which were used in random order: insufflation by intermittent positive-pressure breathing (IPPB) combined with MAC, MI-E and MI-E + MAC.

RESULTS

Overall, PCF was higher with IPPB + MAC than with MI-E + MAC or MI-E alone. Among the 12 patients who had higher PCF values with IPPB + MAC than with the two other techniques, 9 exhibited mask pressure swings during MI-E exsufflation, with a transient positive-pressure value due to the expiratory flow produced by the combined patient cough effort and MAC. Each of these 9 patients had higher PCF values (>5 liters/s) than did the other 9 patients when using IPPB + MAC.

CONCLUSION

Our results indicate that adding the MI-E device to MAC is unhelpful in patients whose PCF with an insufflation technique and MAC exceeds 5 liters/s. This is because the expiratory flow produced by the patient's effort and MAC transitorily exceeds the vacuum capacity of the MI-E device, which therefore becomes a transient load against the PCF.

Для цитирования: Lacombe Matthieu, Del Amo Castrillo Lorena, Boré Aurélien, Chapeau David, Horvat Eric, Vaugier Isabelle, Lejaille Michelle, Orlikowski David, Prigent Hélène, Lofaso Frédéric. Comparison of three cough-augmentation techniques in neuromuscular patients: mechanical insufflation combined with manually assisted cough, insufflation-exsufflation alone and insufflation-exsufflation combined with manually assisted cough.. Respiration; international review of thoracic diseases. 2014. DOI: 10.1159/000364911.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.