Safety and efficacy of atrial fibrillation ablation guided by Ablation Index module.

Solimene Francesco, Schillaci Vincenzo, Shopova Gergana, Urraro Francesco, Arestia Alberto, Iuliano Assunta, Maresca Fabio, Agresta Alessia, La Rocca Vincenzo, De Simone Antonio, Stabile Giuseppe

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Reconnection of pulmonary veins (PVs) remains common following radiofrequency catheter ablation for atrial fibrillation (AF). Ablation Index (AI) is a novel ablation quality marker that incorporates stability, contact force (CF), time, and power in a weighted formula. Its use seems to improve lesion durability. This is a prospective, single-arm registry to investigate on the safety and mid-term efficacy of AF ablation guided by the AI.

METHODS

One hundred fifty-six consecutive patients (mean age 58 ± 10 years, 49% males, 44% with structural heart disease) referred for paroxysmal (124) or persistent (32) AF underwent antral PV isolation using a surround flow CF-sensing catheter guided by the AI. Radiofrequency was delivered targeting interlesion distance ≤ 6 mm and Ablation Index of 330-350 at posterior wall and 400-450 at anterior wall.

RESULTS

Mean overall procedure time was 95 ± 30 min with a mean fluoroscopy time of 5 ± 6 min. Mean ablation time was 26 ± 10 min, 627/628 targeted PV were isolated. One pericardial effusion and two groin hematomas were reported; none required intervention. During a mean follow-up of 14 ± 6 months, 17 (10.8%) (9% paroxysmal AF vs 22% persistent AF, p = 0.09) patients had an atrial arrhythmia recurrence.

CONCLUSIONS

PV ablation guided by AI resulted feasible, achieving a high rate of isolated PVs, with a low complication rate, and allowed a high single-procedure arrhythmia-free survival at 14 months.

Для цитирования: Solimene Francesco, Schillaci Vincenzo, Shopova Gergana, Urraro Francesco, Arestia Alberto, Iuliano Assunta, Maresca Fabio, Agresta Alessia, La Rocca Vincenzo, De Simone Antonio, Stabile Giuseppe. Safety and efficacy of atrial fibrillation ablation guided by Ablation Index module.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2018. DOI: 10.1007/s10840-018-0420-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.