Аннотация

BACKGROUND

Atrioventricular node (AVN) ablation is effective for rate control in atrial fibrillation. This may require multiple radiofrequency applications to achieve complete atrioventricular block (CAB). In this retrospective study, we tested the hypothesis that mapping the AVN utilizing electrograms (EGMs) on both proximal and distal bipoles of the mapping catheter may improve the likelihood of CAB.

METHODS

Lesion characteristics and EGM components on the proximal and distal bipoles of the ablation catheter in first-time AVN ablation procedures were analyzed. Outcomes of each lesion, including presence of CAB, acute recurrence of AVN conduction, new-onset right bundle branch block (RBBB), and junctional escape rhythm, were analyzed. Multivariate binary logistic regression analysis was performed to identify EGM characteristics that independently predicted the outcomes of interest. Lesions with these EGM characteristics were then identified and their outcomes compared with the whole cohort.

RESULTS

A total of 441 ablation lesions were analyzed. EGM characteristics that independently predicted outcomes were the presence of His and atrial EGMs on the distal bipole and the absence of ventricular EGM on the proximal bipole. Among the 25 lesions with all these characteristics, 18 (72%) resulted in CAB compared to the overall cohort rate of 38% (P = 0.001). There was no new-onset RBBB. The likelihood of acute recurrent AVN conduction and junctional escape rhythm were similar.

CONCLUSION

Combining proximal and distal bipole EGM characteristics of the ablation catheter can improve the accuracy of AVN localization during AVN ablation and avoid right bundle branch injury.

Для цитирования: Yeo Wee Tiong, Sporton Simon C E, Dhinoja Mehul, Schilling Richard J, Earley Mark J. Utility of mapping signals to improve precision of atrioventricular node ablation.. Pacing and clinical electrophysiology : PACE. 2013. DOI: 10.1111/pace.12199.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.