MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.otohns.2010.04.267

Медицинская библиотека · Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery · 2010

Long-term results with the Rion E-type semi-implantable hearing aid.

Komori Masahiro, Yanagihara Naoaki, Hinohira Yasuyuki, Hato Naohito, Gyo Kiyofumi

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

The Rion implantable hearing aid (IHA) Ehime (E)-type was developed for ears with middle ear diseases. This study focused on the current status of the patients, device problems, postoperative difficulties, and preventive measures against them.

STUDY

DESIGN

Case series with chart review.

SETTING

Tertiary referral hospital.

SUBJECTS AND

METHODS

Subjects were 30 patients who were implanted with the IHA E-type between 1984 and 1997 and followed up for more than 10 years. Current status of IHA implantees, incidents of device problems, and postoperative troubles and hearing outcomes were reviewed.

RESULTS

Eleven patients (36.7%) still use the original device. The average period of use was 16.6 +/- 3.3 years (21 years at most). The incidence of problems was lower with the second version of the device compared to the first version. Frequencies of the troubles were related to the types of original ear diseases: seven of 17 cases with chronic otitis media (41.2%), two of seven cases with cholesteatoma (28.6%), and two of six cases with tympanosclerosis (33.3%). No cholesteatoma occurred after surgical procedures (i.e., external ear canal closure and tympanic membrane lateralized) (P = 0.06). The device was exposed through a retroauricular skin fistula where the internal coil had been implanted. Significantly fewer infections were observed when the two-stage operation was used (P < 0.01).

CONCLUSION

For long-term success in implantation of the IHA, careful control of middle ear inflammation and measures against eustachian tube dysfunction are required in addition to technological advancements.

Для цитирования: Komori Masahiro, Yanagihara Naoaki, Hinohira Yasuyuki, Hato Naohito, Gyo Kiyofumi. Long-term results with the Rion E-type semi-implantable hearing aid.. Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery. 2010. DOI: 10.1016/j.otohns.2010.04.267.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.