MedElevator Войти в исследования
← Поиск публикаций
Медицинская библиотека · South Dakota medicine : the journal of the South Dakota State Medical Association · 2016

Acute Stent Fracture During Balloon Post Dilation of an Everolimus Eluting Stent.

Yee Jimmy, Kelly Shawn, Kumar Vishesh, Raizada Amol, Anuwatworn Amornpol, Stys Tomasz, Stys Adam

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Coronary artery stent fracture is a well described complication during percutaneous intervention, with rates ranging from 0.84 to 8.4 percent in first generation drug eluting stents. Complications of stent fractures usually present with symptoms of acute coronary syndrome or progressive angina days, months to years after initial implantation. We present a case of an acute stent fracture during post balloon dilation of an everolimus eluting stent at a critical stenosis junction of a saphenous vein graft to the first diagonal of the left anterior descending artery. A shorter oversized drug eluting stent was placed to cover the stent fracture with good angiographic results. To our knowledge, this is the first incidence in literature of an acute stent fracture in a saphenous vein graft.

Для цитирования: Yee Jimmy, Kelly Shawn, Kumar Vishesh, Raizada Amol, Anuwatworn Amornpol, Stys Tomasz, Stys Adam. Acute Stent Fracture During Balloon Post Dilation of an Everolimus Eluting Stent.. South Dakota medicine : the journal of the South Dakota State Medical Association. 2016.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.