Balloon-Expandable TAVR Bioprostheses: Area or Perimeter Sizing? A Prospective Pilot Study.
Аннотация
OBJECTIVE
In TAVR, area sizing is used for balloon-expandable (BE) valves, whereas self-expanding valves are sized to annulus perimeter. For BE valves, this seems illogical: these frames force a circular shape even on an ellipsoid annulus. This can potentially lead to relative undersizing when area sizing is being applied. We developed a perimeter-based sizing algorithm to evaluate the safety and feasibility of perimeter sizing for the Myval BE valve.
METHODS
In this prospective single-center study, 60 patients with severe aortic stenosis treated with the Myval BE valve were included. Perimeter sizing was used with limited oversizing of 3.7% ± 1.3% compared to the annulus perimeter. After TAVR, clinical outcomes were evaluated at 30 days and 1 year. An echocardiographic follow-up took place at 30 days.
RESULTS
At 30 days, the need for PPI and stroke occurred in 2% and 3% of the patients, respectively. Moreover, cardiac death and moderate-severe PVL were absent. At 1-year, cardiac death and stroke were observed in 3% and 8% of the patients, respectively. In 33.3% of the patients, a larger valve size was implanted compared to the valve size calculated by area sizing.
CONCLUSIONS
Perimeter sizing with the Myval BE valve leads to substantial use of larger valve sizes and favorable clinical outcomes, with low PPI and the absence of significant PVL. A randomized controlled trial is being planned to prove the superiority of this alternative sizing method.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.