BACKGROUND

Vascular complications (VCs) after transfemoral transcatheter aortic valve implantation (TAVI) have an increased mortality risk, and vascular closure device (VCD) use is mandatory. The percutaneous MANTA VCD (Teleflex) is a novel collagen-based technology for closure of large-bore arteriotomies. We compared the MANTA VCD with the suture-based ProGlide VCD (Abbott Vascular).

METHODS

A retrospective review was performed on all consecutive patients who underwent transfemoral TAVI in our center from January 1, 2015, to February 28, 2021, and 30-day outcomes were recorded. Major adverse cardiovascular events (MACE) were cardiac death, disabling stroke, and/or major VCs. Access site-related VCs were VCs related to the access site vessel from which the transcatheter valve was introduced and advanced.

RESULTS

The MANTA VCD was used in 99 patients and the ProGlide in 224. There was 4.0% MACE in the MANTA group and 4.9% in the ProGlide group (P = .999). Overall VCs were 10.1% vs 7.6%, major VCs were 3.0% vs 2.2%, and minor VCs 7.1% vs 5.4%(P = .753). Access site-related VCs were 5.1% vs 5.8% in the (P = .999), and periprocedural vascular surgical intervention was needed in 6.1% of the MANTA group vs 2.2% of the ProGlide group (P = .099).

CONCLUSION

There was no significant difference in MACE, mortality, cardiovascular mortality, VCs, access site-related VCs, periprocedural vascular surgical interventions, bleeding, or transfusion rate between the 2 groups. The MANTA VCD group had more periprocedural vascular surgical interventions which did not reach statistical significance.

Для цитирования: Eftychiou Christos, Eteocleous Nicolaos, Mitsis Andreas, Zittis Ioannis, Papadopoulos Kyriacos, Petrou Andria, Adamou Maria, Kounnos Christos, Bekos Christos, Avraamides Panayiotis. Comparison of MANTA vs ProGlide Vascular Closure Device and 30-Day Outcomes in Transfemoral Transcatheter Aortic Valve Implantation.. Texas Heart Institute journal. 2022. DOI: 10.14503/thij-21-7650.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.