Mechanics of lumenless pacing lead strength during extraction procedures based on laboratory bench testing.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Vatterott Pierce J, Mondésert Blandine, Marshall Mark, Lulic Thomas, Wilkoff Bruce L
Аннотация
BACKGROUND: With the advent of conduction system pacing, use of the Medtronic SelectSecure Model 3830 lead has increased substantially. However, with this increased use, the potential need for lead extraction also will increase. Lumenless lead construction requires an understanding of both applicable tensile forces as well as lead preparation techniques that can influence consistent extraction.
OBJECTIVE: The purpose of this study was to use bench testing methodologies to characterize the physical properties of lumenless leads and to describe related lead preparation methods that support known extraction techniques.
METHODS: Multiple 3830 lead preparation techniques, commonly used in extraction practices, were compared on the bench to assess rail strength (RS) in simple traction and use conditions with simulated scar. Retention of the IS1 connector vs severing the lead body preparation techniques were compared. Distal snare and rotational extraction tools were evaluated.
RESULTS: The retained connector method provided higher RS compared to the modified cut lead method: mean 11.42 lbf (9.85-12.73 lbf) vs 8.51 lbf (1.66-14.32 lbf), respectively. Snare use distally did not significantly affect RS: mean 11.05 lbf (8.58-13.95 lbf). Lead damage occurred with the TightRail extraction tool at angles ≥90°, which could occur with right-sided implants.
CONCLUSIONS: When extracting SelectSecure leads, the retained connector method to maintain cable engagement benefits preservation of the extraction RS. Limiting traction force to <10 lbf (4.5 kgf) and avoiding poor lead preparation methods are critical to consistent extraction. Femoral snaring does not change RS when needed and offers a method to regain lead rail in cases of distal cable fracture.
OBJECTIVE: The purpose of this study was to use bench testing methodologies to characterize the physical properties of lumenless leads and to describe related lead preparation methods that support known extraction techniques.
METHODS: Multiple 3830 lead preparation techniques, commonly used in extraction practices, were compared on the bench to assess rail strength (RS) in simple traction and use conditions with simulated scar. Retention of the IS1 connector vs severing the lead body preparation techniques were compared. Distal snare and rotational extraction tools were evaluated.
RESULTS: The retained connector method provided higher RS compared to the modified cut lead method: mean 11.42 lbf (9.85-12.73 lbf) vs 8.51 lbf (1.66-14.32 lbf), respectively. Snare use distally did not significantly affect RS: mean 11.05 lbf (8.58-13.95 lbf). Lead damage occurred with the TightRail extraction tool at angles ≥90°, which could occur with right-sided implants.
CONCLUSIONS: When extracting SelectSecure leads, the retained connector method to maintain cable engagement benefits preservation of the extraction RS. Limiting traction force to <10 lbf (4.5 kgf) and avoiding poor lead preparation methods are critical to consistent extraction. Femoral snaring does not change RS when needed and offers a method to regain lead rail in cases of distal cable fracture.