Suboptimal geometrical implantation of biological aortic valves provokes functional deficits.
Аннотация
Endovascular valves have become a valid option for patients not qualifying for conventional surgery. Biological valves mounted in a stent are currently used. After implantation, however, geometrical distortion of the valve can occur. We tested whether biological valves suitable for transcatheter implantation exhibit hemodynamic deficits after deployment in a distorted position. Two types of valves [bovine pericardium (BP) and porcine cusps], of 21 and 23 mm diameter, respectively were investigated. Mean transvalvular gradient (TVG), effective orifice area (EOA), and regurgitation fraction (REG) were measured prior to and after the 20% distortion of the original diameter. All valves exhibited an increase of TVG and reduction of EOA whereas REG increased only in BP valves after distortion. The 21 mm valves demonstrated a more pronounced alteration than the 23 mm valves. Even moderately distorted implantation of a biological valve results in a marked functional alteration. The susceptibility of pericardial valves is higher than that of porcine valves probably owing to better coaptation properties of native cusps even under deformed conditions when compared to valves constructed with pericardium. Care should therefore be taken during implantation of endovascular valves in order to avoid fixed hemodynamic deficits. Native valves may preferably be used as they demonstrate a more robust behavior regarding suboptimal implantation.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.