Continuous ECG monitoring versus mobile telemetry: A comparison of arrhythmia diagnostics in human- versus algorithmic-dependent systems.

Willcox Mark E, Compton Steven J, Bardy Gust H

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Clinicians rarely scrutinize the full disclosure of a myriad of FDA-approved long-term rhythm monitors, and they rely on manufacturers to detect and report relevant rhythm abnormalities.

OBJECTIVE

The objective of this study is to compare the diagnostic accuracy between mobile cardiac telemetry (MCT), which uses an algorithm-based detection strategy, and continuous long-term electrocardiography (LT-ECG) monitoring, which uses a human-based detection strategy.

METHODS

In an outpatient arrhythmia clinic, we enrolled 50 sequential patients ordered to wear a 30-day MCT, to simultaneously wear a continuous LT-ECG monitor. Periods of concomitant wear of both devices were examined using the associated report, which was over-read by 2 electrophysiologists.

RESULTS

Forty-six of 50 patients wore both monitors simultaneously for an average of 10.3 ± 4.4 days (range: 1.2-14.8 days). During simultaneous recording, patients were more often diagnosed with arrhythmia by LT-ECG compared to MCT (23/46 vs 11/46), P = .018. Similarly, more arrhythmia episodes were detected during simultaneous recording with the LT-ECG compared to MCT (61 vs 19), P < .001. This trend remained consistent across arrhythmia subtypes, including ventricular tachycardia (13 patients by LT-ECG vs 7 by MCT), atrioventricular (AV) block (3 patients by LT-ECG vs 0 by MCT), and AV node reentrant tachycardia (2 patients by LT-ECG vs 0 by MCT). Atrial fibrillation (AF) was documented by both monitors in 2 patients; however, LT-ECG monitoring captured 4 additional AF episodes missed by MCT.

CONCLUSION

In a time-controlled, paired analysis of 2 disparate rhythm monitors worn simultaneously, human-dependent LT-ECG arrhythmia detection significantly outperformed algorithm-based MCT arrhythmia detection.

Для цитирования: Willcox Mark E, Compton Steven J, Bardy Gust H. Continuous ECG monitoring versus mobile telemetry: A comparison of arrhythmia diagnostics in human- versus algorithmic-dependent systems.. Heart rhythm O2. 2021. DOI: 10.1016/j.hroo.2021.09.008.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.