A Digital Electrocardiographic System for Assessing Myocardial Electrical Instability: Principles and Applications.

Vorobiev A P, Vaykhanskaya T G, Melnikova O P, Krupenin V P, Polyakov V B, Frolov A V

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

UNLABELLED: The aim of the study was to develop an ECG hardware and software system for monitoring electrical instability of the myocardium and to assess the diagnostic and prognostic capabilities of this setup in a cardiology clinic.

MATERIALS AND

METHODS

The Intecard 7.3 software and hardware system developed in this study makes it possible to measure fluctuations of the ECG amplitude-time parameters using the beat-to-beat mode. Intecard 7.3 evaluates a number of ECG markers that reflect electrical instability of the myocardium. Among them are the fragmented QRS complex, the spatial QRS-T angle, the T-wave alternans, the duration, and dispersion of the QT interval, the turbulence and acceleration/deceleration of the heart rhythm.Clinical trials of Intecard 7.3 were carried out with 734 patients with ischemic heart disease or cardiomyopathy and 112 healthy individuals.

RESULTS

Intecard 7.3 reliably identifies fragmented QRS complexes by detecting short spikes of <25 ms in the ascending parts of the Q, R, and S waves. The QRS-T angle is determined from the reference amplitudes of the R and T waves in leads avF, V2, V5, and V6. Digital precision processing of the ECG signal improves its accuracy to microvolts and microseconds.The software was designed to measure the T-wave amplitude in each of 300-500 cardiobeats; T-wave alternans was estimated by the moving average method. In a typical cardiobeat, the QT dispersion was calculated based on 12 ECG leads. From the sequence of RR intervals, turbulence, and deceleration of the heart rhythm were determined.During the observation period of 5.0 [2.1; 5.9] years, 90 out of 734 patients (12.3%) experienced adverse cardiovascular events (ACVE). In this period, the myocardial electrical instability was recorded in patients with ACVE more frequently than in those without ACVE. Thus, the frequency of fragmented QRS was 72.2±4.7 vs 16.8±1.5% (p<0.01), the values of the QRS-T angle were 128 [55; 101] vs 80 [53; 121]° (p<0.001), the T-wave alternans - 36.9 [15.5; 62.1] vs 21.9 [10.2; 30.7] μV (p<0.005), the QT interval - 408 [383; 438] vs 376 [351; 400] ms (p<0.001), the QT dispersion - 76 [57; 96] vs 64 [50; 92] ms (p<0.005), respectively. In patients with ACVE, the threshold that triggers pathological rhythm turbulence was higher (>0%) than that in healthy controls (p<0.001); the deceleration of the heart rhythm was reduced from 19.2 [2.2; 38.0] to 8.8 [4.0; 16.8] ms (p<0.05).A personalized model for ACVE risk stratification has been developed. In this model, the area under the ROC curve was 0.856; sensitivity - 75%; specificity - 78%; predictive accuracy - 77%.

CONCLUSION

Using the ECG markers of myocardial electrical instability, the Intecard 7.3 system allows one to predict life-threatening ventricular tachyarrhythmias and sudden cardiac death with an accuracy of 77%. The non-invasiveness, high productivity, and reasonable cost ensure the availability of this predictive technology in all levels of healthcare.

Для цитирования: Vorobiev A P, Vaykhanskaya T G, Melnikova O P, Krupenin V P, Polyakov V B, Frolov A V. A Digital Electrocardiographic System for Assessing Myocardial Electrical Instability: Principles and Applications.. Sovremennye tekhnologii v meditsine. 2020. DOI: 10.17691/stm2020.12.6.02.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.