A case report of lead dysfunction presenting as high ventricular premature complex burden.
СтатьяПеревод пока не подготовлен
Аннотация
Pacemaker-troubleshooting is an important step in the evaluation of a patient with syncope post-pacemaker-implantation. The basic functions of sensing, pacing and impedance may remain spuriously normal in the case of lead-microfracture or insulation break. We report a case in which the lead dysfunction was diagnosed based on multiple episodes of premature ventricular beats.
Для цитирования: Ajit Deshpande Saurabh, Udyavar Ameya. A case report of lead dysfunction presenting as high ventricular premature complex burden.. Indian pacing and electrophysiology journal. 2021. DOI: 10.1016/j.ipej.2021.10.003.