Preoperative assessment of the carotid bifurcation. Can magnetic resonance angiography and duplex ultrasonography replace contrast arteriography?

Patel M R, Kuntz K M, Klufas R A, Kim D, Kramer J, Polak J F, Skillman J J, Whittemore A D, Edelman R R, Kent K C

Клиническое исследованиеСравнительное исследованиеКонтролируемое клиническое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND PURPOSE: Noninvasive studies are used with increasing frequency to assess the carotid bifurcation before endarterectomy. Therefore, assessment of their diagnostic accuracies is essential for appropriate patient management. We prospectively evaluate two noninvasive tests, magnetic resonance angiography (MRA) and duplex ultrasonography (DU), as potential replacements for contrast arteriography (CA).

METHODS

A blinded comparison of three-dimensional time-of-flight (TOF) MRA, two-dimensional TOF MRA, and DU in 176 arteries was performed. CA was used as the standard of comparison.

RESULTS

Three-dimensional TOF MRA had a sensitivity of 94%, a specificity of 85%, and an accuracy of 88% for the identification of 70% to 99% stenosis; two-dimensional TOF MRA had a sensitivity and specificity that were approximately 10% lower than those of three-dimensional TOF MRA. DU resulted in a sensitivity of 94%, a specificity of 83%, and an accuracy of 86%. Combining data from three-dimensional TOF MRA and DU, allowing for CA only for disparate results, yielded a sensitivity of 100%, a specificity of 91%, and an accuracy of 94% among concordant noninvasive tests, with CA required in 16% of arteries. MRA accurately differentiated 17 carotid occlusions from 16 high-grade (90% to 99%) stenoses, whereas with DU two patent arteries were identified as occluded and one occluded artery was identified as patent.

CONCLUSIONS

Three-dimensional TOF MRA is the most accurate noninvasive test. Combined use of MRA and DU results in a marked increase in accuracy to a level that obviates the need for CA in a majority of patients.

Для цитирования: Patel M R, Kuntz K M, Klufas R A, Kim D, Kramer J, Polak J F, Skillman J J, Whittemore A D, Edelman R R, Kent K C. Preoperative assessment of the carotid bifurcation. Can magnetic resonance angiography and duplex ultrasonography replace contrast arteriography?. Stroke. 1995. DOI: 10.1161/01.str.26.10.1753.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.