Clinical Experience of Vibroplasty With Direct Coupling to the Oval Window Without Use of a Coupler.
Аннотация
OBJECTIVES/HYPOTHESIS: To investigate the efficacy of direct implantation of a Vibrant Soundbridge (VSB) implant in the oval window (OW) without the use of an OW coupler in patients with severe mixed hearing loss.
STUDY
DESIGN
Retrospective chart review
METHODS
A total of 62 patients underwent VSB implantation between July 2016 and December 2018 at Severance Hospital in Seoul, South Korea. Among them, eight patients (nine ears) with moderate-to-severe mixed hearing loss were implanted with a VSB directly in the OW. A floating mass transducer (FMT) was attached to the stapes footplate and covered with tragal cartilage. The outcomes were evaluated using pure-tone audiogram and speech audiogram preoperatively and postoperatively. Word recognition score (WRS; % correct) were measured at the most comfortable loudness (MCL) level to evaluate speech perception.
RESULTS
All cases posed difficulty with round window vibroplasty during surgery, and eventually, an FMT was appropriately placed in the OW without a coupler. Preoperative and postoperative bone conduction thresholds were not different. VSB-aided threshold improved in terms of functional and effective gains. Interestingly, four cases showed improved air conduction thresholds without the use of a VSB. In addition, MCL level with a VSB was significantly lower than that with a hearing aid, and VSB-aided WRS improved over time.
CONCLUSIONS
Direct implantation of a VSB in the OW without the use of a coupler showed favorable hearing outcomes, and the OW vibroplasty was safe. Direct OW vibroplasty without a coupler is a reliable procedure and can be a good option for hearing rehabilitation in patients with severe mixed hearing loss.
LEVEL OF EVIDENCE: 4 Laryngoscope, 2020.