Медицинская библиотека · Clinical case reports · 2026

When Stability Fails: Rare Very Late Dislodgement of 3830 Lead.

Kumar Dilip, Ghosh Niladri, Halder Ashesh, Chakraborty Rabin, Kulshrestha Srashti, De Arnab

СтатьяПеревод в очереди

Аннотация

We report a rare case of very late dislodgement of a Medtronic SelectSecure 3830 lumenless lead three years after successful left bundle branch area pacing. An 88-year-old patient with permanent atrial fibrillation and bifascicular block underwent VVIR pacemaker implantation in 2022 with excellent initial parameters (threshold 0.8 V at 0.5 ms, impedance 650 Ω) and stable follow-up for three years. The patient presented with pre-syncopal episodes, and device interrogation revealed capture failure with elevated threshold (2.25 V/0.4 ms) and decreased impedance. Fluoroscopic comparison confirmed macro-dislodgement of the lead with coiling in the pacemaker pocket. The SelectSecure lead was successfully extracted and replaced with a conventional stylet-driven lead positioned in the right ventricular mid-septum. This case highlights that late lead dislodgement can occur with lumenless leads even years after successful implantation, emphasizing the need for continued vigilance during long-term follow-up of cardiac physiological pacing systems.

Для цитирования: Kumar Dilip, Ghosh Niladri, Halder Ashesh, Chakraborty Rabin, Kulshrestha Srashti, De Arnab. When Stability Fails: Rare Very Late Dislodgement of 3830 Lead.. Clinical case reports. 2026. DOI: 10.1002/ccr3.71819.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.