Feasibility and safety of endomyocardial biopsy by lumenless pacing lead sheath-directed method during cardiac implantable electronic device implantation.

Ye Yang, Chen Youyou, Jiang Hao, Zeng Jiling, Wang Xingchen, Yang Ying, Liu Xiang Lan, Hong XuLin, Wang Bei, Sun Ya Xun, Lai Dongwu, Xu Xi-Qi, Fu Guo-Sheng

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Patients with an indication for a cardiovascular implantable electronic device (CIED) are complicated with special cardiomyopathy or other unspecified cardiac abnormalities and may need endomyocardial biopsy (EMB). However, EMB by a bioptome is usually avoided to reduce the risk of lead displacement in the CIED periprocedural period.

OBJECTIVE

We aimed to assess the safety and feasibility of a novel approach for transvenous right ventricular (RV) EMB using the lead sheath method (L-S-M) during CIED implantation and compared it with the traditional bioptome method (T-B-M).

METHODS

A consecutive 80 patients referred for EMB were enrolled. In the L-S-M group, a sheath with a lumenless pacing lead was positioned toward the middle to apical interventricular septum under fluoroscopy (n = 60). The CIED implantation was performed through the same venous access site. In the T-B-M group, a bioptome was used (n = 20). The clinical characteristics, procedural details, success rate, and complications were evaluated.

RESULTS

There were 380 RV EMB procedures performed with comparable numbers of samples per patient, 4.1 ± 0.8 in the L-S-M group and 3.8 ± 0.8 in the T-B-M group. In the L-S-M group, 7 (11.7%) patients experienced minor complications with 3 transient right bundle branch blocks, 2 transient atrioventricular blocks, and 2 regional minor device pocket hematomas. In the T-B-M group, 3 (15%) experienced 1 temporary pacing for transient atrioventricular block, 1 chest discomfort, and 1 regional hematoma. No cardiac tamponade was detected.

CONCLUSION

RV EMB by the innovative L-S-M is technically feasible and safe. It can yield valuable and early diagnostic insights for patients who are candidates for CIEDs.

Для цитирования: Ye Yang, Chen Youyou, Jiang Hao, Zeng Jiling, Wang Xingchen, Yang Ying, Liu Xiang Lan, Hong XuLin, Wang Bei, Sun Ya Xun, Lai Dongwu, Xu Xi-Qi, Fu Guo-Sheng. Feasibility and safety of endomyocardial biopsy by lumenless pacing lead sheath-directed method during cardiac implantable electronic device implantation.. Heart rhythm. 2024. DOI: 10.1016/j.hrthm.2024.11.018.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.