MedElevator Войти в исследования
← Поиск публикаций

DOI 10.5301/ijao.5000626

Медицинская библиотека · The International journal of artificial organs · 2017

Thrombosis in left ventricular assistance device with centrifugal technology: is early thrombolysis a better solution?

Centofanti Paolo, Baronetto Andrea, Attisani Matteo, Ricci Davide, Simonato Erika, La Torre Michele W, Boffini Massimo, Rinaldi Mauro

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Continuous flow left ventricular assistance devices (CF-LVADs) have revolutionized the treatment of advanced heart failure. Pump replacement for thrombosis is a high-risk procedure with a high perioperative mortality rate with possible recurrence. We aim to summarize our experience using a conservative approach with medical therapy.

METHODS

We retrospectively reviewed records of patients who experienced pump thrombosis after LVAD implantation with HeartWare HVAD at our institution, from November 2010 to March 2016. Device thrombosis (DT) was divided into suspected (SDT) and confirmed (CDT). A conservative approach using thrombolysis and heparin was used in all patients.

RESULTS

A total of 32 HeartWare HVAD pumps were implanted. Mean age was 59 ± 10 years and the mean time on mechanical support was 19.29 months (±14.06). Pump thrombosis occurred in 7 patients (0.14 patients/year) after a mean time of 733 (231-1,606) days after LVAD implantation. Three out of 7 cases had thrombosis recurrence (43%). Overall 19 episodes were recorded (0.38 event per patient/year). Eighteen out of 19 thrombolytic treatments were successful (94.7%). No patient required LVAD replacement or transfusion of blood products. There was no significant difference in terms of survival between patients who experienced thrombotic events and patients who did not. No major complications related to thrombolysis were recorded.

CONCLUSIONS

Systemic thrombolysis plus heparin was an excellent therapeutic option. Early intervention in clinically stable patients without signs of heart failure but with indirect signs of device thrombosis has led to better outcomes.

Для цитирования: Centofanti Paolo, Baronetto Andrea, Attisani Matteo, Ricci Davide, Simonato Erika, La Torre Michele W, Boffini Massimo, Rinaldi Mauro. Thrombosis in left ventricular assistance device with centrifugal technology: is early thrombolysis a better solution?. The International journal of artificial organs. 2017. DOI: 10.5301/ijao.5000626.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.