Comparison of the accuracy of an ultrasonic-based jaw tracking device with conventional electronic tracking device.
Аннотация
PURPOSE: This study aimed to evaluate the accuracy of the ultrasonic jaw tracking system by comparing with the conventional electronic system in recording condylar movements.
MATERIALS AND
METHODS
Twenty-six subjects with normal occlusion participated in the study. The CADIAX® 4 and Jaw Motion Analyzer (JMA) systems were used to record condylar movement trajectories during mandibular border movements (protrusive/retrusive, lateral, and wide mouth opening), with each movement repeated three times. Both systems used facebows and sensors to locate the condylar axis points and capture movement trajectory data. Paired t-tests were used for normally distributed data, while the Wilcoxon rank-sum test was applied to non-normally distributed data. The level of significance was set at α = .05.
RESULTS
The maximum condylar displacement in the sagittal plane during mandibular border movements and the sagittal condylar inclination (SCI) values on both the left and right sides showed no significant difference between the two systems (P > .05). The Bennett angle (BA) values on both the left and right sides measured by the JMA system were significantly higher than those measured by the CADIAX® 4 system (P < .05). The comfort levels of the JMA system were significantly higher than the CADIAX® 4 system (P < .05).
CONCLUSION
Through this study, it was found that the accuracy of the ultrasonic jaw tracking system was comparable with the conventional electronic system, except for the Bennett angle measurement. In terms of comfort and ease of use, the ultrasonic jaw tracking system is more favored.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.