DOI 10.1136/heartjnl-2018-314186
External continuous ECG versus loop recording for atrial fibrillation detection in patients who had a stroke.
Аннотация
BACKGROUND
Detection of atrial fibrillation (AF) in patients who had ischaemic stroke and transient ischaemic attack (IS/TIA) is recommended. We aimed to compare external loop recording (ELR) against simultaneous continuous ECG recording for AF detection in patients who had acute IS/TIA and determine sensitivity, specificity and positive predictive value of AF detection using ELR. We hypothesised ELR to detect 15% fewer patients with AF than continuous ECG recording.
METHODS
In this prospective cohort study, we included 1412 patients who had acute IS/TIA without prior AF. Monitoring was 48 hours. Primary outcome was AF >30 s. Cardiologist verified AF in continuous ECG was gold standard.
RESULTS
In continuous ECG, 38 (2.7%) patients had AF. ELR automatically categorised 219/1412 patients (15.5%) with AF, including 32/38 (85%) patients with AF in continuous ECG. After cardiologist adjudication of ELR recordings, AF was diagnosed in 57/219 patients, of which 32 (56%) had AF in continuous ECG. For adjudicated AF detection by ELR, sensitivity was 84%, 95% CI (69% to 94%), specificity was 98%, 95% CI (97% to 99%) and positive predictive value was 56%, 95% CI (42% to 69%).
CONCLUSION
Automatic AF detection with ELR results in an AF diagnosis in more than five patients without AF for each patient with AF as verified in continuous ECG. For adjudicated AF detection by ELR, sensitivity was confirmed to 84% and specificity 98%. Automatic ELR as investigated in this study may be considered to rule out AF, but it is not suitable as a single monitoring device for AF screening in patients early after stroke.
TRIAL REGISTRATION NUMBER: NCT02155907.