DOI 10.1097/rct.0000000000001228
Magnetic Resonance Elastography of the Liver: A Single-Institution Review of Cases Performed in a 5-Year Interval.
Аннотация
OBJECTIVE
This study aimed to assess the diagnostic accuracy of magnetic resonance elastography (MRE) in detecting hepatic fibrosis and determining clinically relevant stiffness cutoff values per stage of fibrosis.
METHODS
This retrospective study assessed 1488 hepatic MRE evaluations performed at a single institution for 5 years. Mean liver stiffness measurements were collected from 282 patients who had an MRE study within 1 year of histopathologic analysis. Areas under receiver operating characteristic curves were calculated for each stage of fibrosis with nonparametric ordinal measures of accuracy, and Youden Index was determined.
RESULTS
Mean liver stiffness measurement values were as follows: F0, 2.5± 0.55 kPa; F1, 3.1± 0.80 kPa; F2, 3.4±0.95 kPa; F3, 4.7±1.44 kPa; and F4, 7.9± 2.64 kPa. Nonparametric ordinal measures of accuracy per fibrosis stage were as follows: F0: 0.934, P < 0.001; F0-F1: 0.917, P < 0.001; F0-F2: 0.944, P < 0.001; and F0-F3: 0.941, P < 0.001. Youden Index values for fibrosis stages F2, F3, and F4 were 3.9, 4.0, and 4.5 kPa, respectively.
CONCLUSIONS
Magnetic resonance elastography is an accurate diagnostic tool in assessing liver fibrosis.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.