BACKGROUND AND PURPOSE: Time-resolved (TR) MR angiography (MRA) using parallel imaging techniques is proving to have clinical utility for improving MRA spatial and temporal resolution and separating arterial from venous anatomy. The purpose of this study was to evaluate TR MRA of the intracranial vessels at different integrated parallel acquisition technique (IPAT) factors.
MATERIALS AND
METHODS
3D TR MRA using time-resolved echo-shared angiographic technique was performed with different IPAT factors (0, 2, 3) at 1.5 T, resulting in temporal resolutions of 4.0, 1.7, and 1.3 seconds, respectively. We studied 14 subjects, comprising 12 patients with various pathologic conditions and 2 healthy subjects. The brain volume was covered by 36 partitions, and a bolus of 5 mL of gadopentate dimeglumine was administered. Signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), the number of frames that distinguished between arterial and venous phases, the conspicuity of the vasculature, and artifacts were analyzed.
RESULTS
There was no significant difference in SNR between IPAT factors 0 and 2. Moreover, SNR was significantly lower with IPAT 3 than with IPAT 0 or 2. Smaller vessel segments (M3 and P3) were rated significantly inferior with TR MRA IPAT 2 or 3 compared with MRA without IPAT. For larger proximal vessels (A1 and A2 segments of anterior cerebral artery, M1 and M2 segments of middle cerebral artery, P2 segment of posterior cerebral artery, and basilar artery), there was no difference between TR MRA IPAT 0 and 2.
CONCLUSION
To obtain arterial and venous information in a clinical setting, intracranial TR MRA is best performed with an IPAT factor of 2 with at least 5 mL of contrast.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.