DOI 10.1016/j.brachy.2022.12.008

Медицинская библиотека · Brachytherapy · 2023

Validation of automated post-adjustments of HDR prostate brachytherapy treatment plans by quantitative measures and oncologist observer study.

Dohlmar Frida, Morén Björn, Sandborg Michael, Smedby Örjan, Valdman Alexander, Larsson Torbjörn, Carlsson Tedgren Åsa

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: The aim was to evaluate a postprocessing optimization algorithm's ability to improve the spatial properties of a clinical treatment plan while preserving the target coverage and the dose to the organs at risk. The goal was to obtain a more homogenous treatment plan, minimizing the need for manual adjustments after inverse treatment planning.

MATERIALS AND

METHODS

The study included 25 previously treated prostate cancer patients. The treatment plans were evaluated on dose-volume histogram parameters established clinical and quantitative measures of the high dose volumes. The volumes of the four largest hot spots were compared and complemented with a human observer study with visual grading by eight oncologists. Statistical analysis was done using ordinal logistic regression. Weighted kappa and Fleiss' kappa were used to evaluate intra- and interobserver reliability.

RESULTS

The quantitative analysis showed that there was no change in planning target volume (PTV) coverage and dose to the rectum. There were significant improvements for the adjusted treatment plan in: V150% and V200% for PTV, dose to urethra, conformal index, and dose nonhomogeneity ratio. The three largest hot spots for the adjusted treatment plan were significantly smaller compared to the clinical treatment plan. The observers preferred the adjusted treatment plan in 132 cases and the clinical in 83 cases. The observers preferred the adjusted treatment plan on homogeneity and organs at risk but preferred the clinical plan on PTV coverage.

CONCLUSIONS

Quantitative analysis showed that the postadjustment optimization tool could improve the spatial properties of the treatment plans while maintaining the target coverage.

Для цитирования: Dohlmar Frida, Morén Björn, Sandborg Michael, Smedby Örjan, Valdman Alexander, Larsson Torbjörn, Carlsson Tedgren Åsa. Validation of automated post-adjustments of HDR prostate brachytherapy treatment plans by quantitative measures and oncologist observer study.. Brachytherapy. 2023. DOI: 10.1016/j.brachy.2022.12.008.