MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcv.2020.104650

Медицинская библиотека · Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology · 2020

Pushing beyond specifications: Evaluation of linearity and clinical performance of the cobas 6800/8800 SARS-CoV-2 RT-PCR assay for reliable quantification in blood and other materials outside recommendations.

Nörz Dominik, Frontzek André, Eigner Ulrich, Oestereich Lisa, Wichmann Dominic, Kluge Stefan, Fischer Nicole, Aepfelbacher Martin, Pfefferle Susanne, Lütgehetmann Marc

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The ongoing SARS-CoV-2 pandemic presents a unique challenge to diagnostic laboratories. There are preliminary studies correlating qRT-PCR results from different materials to clinical outcomes, yet, comparability is limited due to the plethora of different assays used for diagnostics. In this study we evaluate clinical performance and linear range for the SARS-CoV-2 IVD (cobas6800/8800 system, a fully automated sample-to-result platform) in different clinically relevant matrix materials outside official specifications.

METHODS

Assay performance was assessed in human plasma, BAL/BL and transport medium following chemical inactivation. For analytical evaluation, respective matrix materials were spiked with SARS-CoV-2 RNA in ten-fold dilution series. The efficacy of chemical inactivation by guanidine hydrochloride solution was confirmed in cell culture infectivity experiments. For correlation, a total of 289 predetermined clinical samples including respiratory swabs, plasma and lower respiratory tract specimens were subjected to the SARS-CoV-2 IVD test and results were compared.

RESULTS

The SARS-CoV-2 IVD showed excellent linearity over four to six log steps depending on matrix material. Chemical inactivation resulted in a reduction in plaque forming units of at least 3.5 log steps, while having no significant impact on assay performance. Inter-run consistency from three different testing sites demonstrated excellent comparability of RT-PCR results (maximum deviation was 1.53 CT). Clinical evaluation for respiratory swabs showed very good agreement with the comparator assay (Positive agreement 95.7 %, negative agreement 98.9 %).

CONCLUSION

The SARS-CoV-2 IVD test for the cobas6800/8800 systems offers excellent linear range and inter-run consistency for quantification of SARS-CoV-2 RNA in different matrices outside official specifications.

Для цитирования: Nörz Dominik, Frontzek André, Eigner Ulrich, Oestereich Lisa, Wichmann Dominic, Kluge Stefan, Fischer Nicole, Aepfelbacher Martin, Pfefferle Susanne, Lütgehetmann Marc. Pushing beyond specifications: Evaluation of linearity and clinical performance of the cobas 6800/8800 SARS-CoV-2 RT-PCR assay for reliable quantification in blood and other materials outside recommendations.. Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology. 2020. DOI: 10.1016/j.jcv.2020.104650.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.