Pooled testing for COVID-19 diagnosis by real-time RT-PCR: A multi-site comparative evaluation of 5- & 10-sample pooling.
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Praharaj Ira, Jain Amita, Singh Mini, Balakrishnan Anukumar, Dhodapkar Rahul, Borkakoty Biswajyoti, Ashok Munivenkatappa, Das Pradeep, Biswas Debasis, Kalawat Usha, Turuk Jyotirmayee, Sugunan A P, Prakash Shantanu, Singh Anirudh K, Barathidasan Rajamani, Subhadra Subhra, Sabat Jyotsnamayee, Manjunath M J, Kanta Poonam, Mudhigeti Nagaraja, Hazarika Rahul, Mishra Hricha, Abhishek Kumar, Santhalembi C, Dikhit Manas Ranjan, Vijay Neetu, Narayan Jitendra, Kaur Harmanmeet, Giri Sidhartha, Gupta Nivedita
Аннотация
METHODS: Ten virus research and diagnostic laboratories (VRDLs) testing for COVID-19 by real-time RT-PCR participated in this evaluation. At each laboratory, 100 nasopharyngeal swab samples including 10 positive samples were used to create 5- and 10-sample pools with one positive sample in each pool. RNA extraction and real-time RT-PCR for SARS-CoV-2-specific E gene target were performed for individual positive samples as well as pooled samples. Concordance between individual sample testing and testing in the 5- or 10-sample pools was calculated, and the variation across sites and by sample cycle threshold (Ct) values was analyzed.
RESULTS: A total of 110 each of 5- and 10-sample pools were evaluated. Concordance between the 5-sample pool and individual sample testing was 100 per cent in the Ct value ≤30 cycles and 95.5 per cent for Ctvalues ≤33 cycles. Overall concordance between the 5-sample pooled and individual sample testing was 88 per cent while that between 10-sample pool and individual sample testing was 66 per cent. Although the concordance rates for both the 5- and 10-sample pooled testing varied across laboratories, yet for samples with Ct values ≤33 cycles, the concordance was ≥90 per cent across all laboratories for the 5-sample pools.
INTERPRETATION & CONCLUSIONS: Results from this multi-site assessment suggest that pooling five samples for SARS-CoV-2 detection by real-time RT-PCR may be an acceptable strategy without much loss of sensitivity even for low viral loads, while with 10-sample pools, there may be considerably higher numbers of false negatives. However, testing laboratories should perform validations with the specific RNA extraction and RT-PCR kits in use at their centres before initiating pooled testing.