Validation study of a portable monitoring device for identifying OSA in a symptomatic patient population.

To Kin-Wang, Chan Wing-Chi, Chan Tat-On, Tung Alvin, Ngai Jenny, Ng Susanna, Choo Kah-Lin, Hui David S

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND AND

OBJECTIVE

Obstructive sleep apnoea syndrome (OSAS) is a common disorder associated with early atherosclerosis, diabetes mellitus, ischaemic heart disease and cerebrovascular disease. The gold standard for confirming OSAS is based on an attended overnight polysomnography (PSG) in a sleep laboratory; however lack of health-care resources creates long waiting times for patient access to this diagnostic test. This study evaluated the ability of a portable sleep-monitoring device to identify patients in Hong Kong with suspected OSAS.

METHODS

Patients with symptoms of OSAS were invited to use the ARES (apnoea risk evaluation system) concurrently with an attended inpatient PSG. Several sets of AHI were generated by the ARES provider based on different oxygen desaturation criteria and surrogate parameters of arousal. The results were compared against PSG to determine the optimal sensitivity and specificity.

RESULTS

There were 141 patients who completed the study successfully. Results of AHI from the ARES study were presented in the order of different scoring criteria--4% oxygen desaturation alone, obstructive events with 3% oxygen desaturation and obstructive events with 1% desaturation plus surrogate arousal criteria. The sensitivity was 0.84 (95% confidence interval (CI): 0.77-0.90), 0.89 (95% CI: 0.89-0.94) and 0.97 (95% CI: 0.94-0.99), respectively. The specificity was 1, 1 and 0.63 (95% CI: 0.55-0.71), respectively. The receiver operating curve had an area of 0.96, 0.97 and 0.98, respectively. The kappa coefficient varied from 0.24 to 0.55 for agreement of severity between PSG and ARES. The likelihood ratio positive and the likelihood ratio negative were 2.61, infinity, infinity and 0.16, 0.11, 0.05, respectively, in the order of oxygen desaturation described earlier.

CONCLUSIONS

The ARES device has reasonable sensitivity and specificity for diagnosing severe OSAS in symptomatic Chinese patients. There is moderate agreement between ARES and PSG in the diagnosis of severe disease, but less agreement in patients with mild/moderate disease.

Для цитирования: To Kin-Wang, Chan Wing-Chi, Chan Tat-On, Tung Alvin, Ngai Jenny, Ng Susanna, Choo Kah-Lin, Hui David S. Validation study of a portable monitoring device for identifying OSA in a symptomatic patient population.. Respirology (Carlton, Vic.). 2008. DOI: 10.1111/j.1440-1843.2008.01439.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.