Novel preclinical method for evaluating the efficacy of a percutaneous treatment in human ex vivo calcified plaque.
Аннотация
The lack of suitable atherosclerotic calcification models and testing strategies inhibits preclinical efficacy testing of existing and novel percutaneous devices. The goal of this study is to develop a preclinical testing method for quantitatively and qualitatively evaluating the efficacy of noncompliant balloon angioplasty (NC BA) treatment in human ex vivo calcified plaque (CP). NC BA using a 3- and 4-mm diameter balloon was performed on an ex vivo tibial calcified vessel obtained from an amputation. Three-dimensional microcomputed tomography (μ-CT) imaging was performed pre- and post-BA to compare crack density in the CP. Comparing the pre- and posttreatment three-dimensional μ-CT images showed a glass-like cracking that occurred in the CP due to the BA procedure. Expansion of the 3-mm balloon showed little tissue deformation and no CP cracking. Although expansion of the 4-mm balloon occurred nonuniformly along balloon length and across the perpendicular projections, the balloon generated cracking throughout the CP, which allowed the surrounding elastic tissue to be dilated. This combined X-ray microscopy and μ-CT technique is a useful preclinical tool for quantifying the efficacy of percutaneous treatments for CP. Because of its nondestructive nature, the CP structure can be visualized pre- and posttreatment to determine the treatment effect.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.