DOI 10.3390/diagnostics14222568
Evaluating Factors Affecting Mean Glandular Dose in Mammography: Insights from a Retrospective Study in Dubai.
Аннотация
BACKGROUND/
OBJECTIVE
This study evaluates the mean glandular dose (MGD) in mammography screening for women aged 40-69 in Dubai, based on a retrospective analysis of a dose survey involving 2599 participants.
METHODS
MGD was calculated using the Dance formula.
RESULTS
The average MGD was 0.96 ± 0.39 mGy for mediolateral oblique (MLO) views and 0.81 ± 0.33 mGy for craniocaudal (CC) views. Weak inverse correlations were found between age and organ dose (OD) for both views, while a direct relationship was observed between breast thickness and entrance skin dose (ESD). In adjusted models, ESD was strongly associated with MGD (β = 1.04, 95% CI: 0.97, 1.09), while OD showed a moderate association (β = 0.44, 95% CI: 0.40, 0.49). Significant variations in ESD, OD, and MGD were noted across age groups and breast thicknesses.
CONCLUSIONS
Lower MGD indicates reduced radiation exposure risk, while higher MGD in MLO views suggests improved imaging quality. Monitoring and optimizing MGD are essential for enhancing patient safety and screening efficacy.