Photon and Neutron Dose Estimation Using Monte Carlo Simulation in TrueBeam's Room.
Аннотация
PURPOSE: The distribution of neutron ambient dose equivalent within the TrueBeam 10 MV photon chamber was investigated.
MATERIALS AND
METHODS
The research used particle and heavy ion transport code system (PHITS) code and JENDL-5.0 to simulate the neutron ambient dose equivalent on and around TrueBeam's head. The simulated results were compared with the measured results using CR-39 detectors when TrueBeam radiated 5000 monitor units of 10 MV photons with field sizes 20 cm × 20 cm and 0.5 cm × 0.5 cm.
RESULTS
Out of field size, the neutron ambient dose equivalents of the 0.5 cm × 0.5 cm field size are higher than those values of the 20 cm × 20 cm field size from 4% to 30%. The differences between the simulated value and the measured value of the neutron ambient dose equivalents at all points out of field size are smaller than 20%.
CONCLUSION
The neutron ambient dose equivalents, simulated with PHITS and JENDL-5.0, are satisfied with the measured neutron ambient dose equivalent.