DOI 10.4103/jmp.jmp_70_24

Медицинская библиотека · Journal of medical physics · 2024

Photon and Neutron Dose Estimation Using Monte Carlo Simulation in TrueBeam's Room.

Quoc Soai Dang, Fujibuchi Toshioh, Arakawa Hiroyuki, Hamada Keisuke

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The distribution of neutron ambient dose equivalent within the TrueBeam 10 MV photon chamber was investigated.

MATERIALS AND

METHODS

The research used particle and heavy ion transport code system (PHITS) code and JENDL-5.0 to simulate the neutron ambient dose equivalent on and around TrueBeam's head. The simulated results were compared with the measured results using CR-39 detectors when TrueBeam radiated 5000 monitor units of 10 MV photons with field sizes 20 cm × 20 cm and 0.5 cm × 0.5 cm.

RESULTS

Out of field size, the neutron ambient dose equivalents of the 0.5 cm × 0.5 cm field size are higher than those values of the 20 cm × 20 cm field size from 4% to 30%. The differences between the simulated value and the measured value of the neutron ambient dose equivalents at all points out of field size are smaller than 20%.

CONCLUSION

The neutron ambient dose equivalents, simulated with PHITS and JENDL-5.0, are satisfied with the measured neutron ambient dose equivalent.

Для цитирования: Quoc Soai Dang, Fujibuchi Toshioh, Arakawa Hiroyuki, Hamada Keisuke. Photon and Neutron Dose Estimation Using Monte Carlo Simulation in TrueBeam's Room.. Journal of medical physics. 2024. DOI: 10.4103/jmp.jmp_70_24.