MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11547-024-01906-8

Медицинская библиотека · La Radiologia medica · 2024

Left radial vs right femoral: comparison between arterial accesses in c-TACE procedures in terms of operator radiations exposure and patient comfort.

Giurazza Francesco, Vizzuso Antonio, Capussela Tiziana, De Martino Fortuna, Pirozzi Palmese Valentina, Giorgetti Gianluigi, Corvino Fabio, Ierardi Anna Maria, Biondetti Pierpaolo, Lucatelli Pierleone, Giampalma Emanuela, Carrafiello Gianpaolo, Niola Raffaella

СтатьяСравнительное исследованиеРандомизированное исследованиеМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: This multicenter prospective study aims to compare transradial access versus transfemoral access in conventional transarterial chemoembolization (c-TACE) procedures, focusing on operators radiations exposure, patients comfort, technical success and vascular access complications.

MATERIALS AND

METHODS

Patients were affected by hepatocellularcarcinoma (HCC) in intermediate stage or previous stages unfit for ablation and/or surgery; they were randomized into two groups according to arterial access site (Group F: right femoral access in standard position; Group R: radial access with left arm abduced 90°). Overall fluoroscopy time was recorded. Eight thermoluminescence dosimeters were positioned immediately before each procedure to monitor radiation doses. Technical success was intended as complete HCC nodules lipiodolization at final plain cone-beam CT.

RESULTS

Group F included 23 patients, while group R 19. Mean fluoroscopy time was lower in group F but difference was not statistically significant (p-value > 0.05). In terms of operators radiations exposure, no significant differences were found (p-value > 0.05). Technical success was obtained in 81.5% in group F and 84.8% in group R, without significant differences (p-value > 0.05). Patients discomfort was significantly (p-value < 0.05) higher in group F. Concerning minor complications, no statistical differences were appreciated (p-value > 0.05); no major complications occurred.

CONCLUSIONS

In this study, no statistical differences were observed in terms of operators radiations exposure, fluoroscopy time and technical success during c-TACE performed with left radial access compared to right femoral access; patients comfort was significantly better with radial access. These data should lead interventional radiologists to favor radial access in c-TACE interventions.

Для цитирования: Giurazza Francesco, Vizzuso Antonio, Capussela Tiziana, De Martino Fortuna, Pirozzi Palmese Valentina, Giorgetti Gianluigi, Corvino Fabio, Ierardi Anna Maria, Biondetti Pierpaolo, Lucatelli Pierleone, Giampalma Emanuela, Carrafiello Gianpaolo, Niola Raffaella. Left radial vs right femoral: comparison between arterial accesses in c-TACE procedures in terms of operator radiations exposure and patient comfort.. La Radiologia medica. 2024. DOI: 10.1007/s11547-024-01906-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.