PURPOSE: Acute pericarditis is a minor complication following atrial fibrillation (AF) ablation procedures. The aim of the study was to evaluate the incidence and clinical aspects of pericarditis following cryoballoon (CB) ablation of AF investigating a possible association with procedural characteristics and a possible relationship with post-ablation recurrences.

METHODS

Four hundred fifty consecutive patients (male 73%, age 59.9 ± 11.2 years) with drug-resistant paroxysmal AF who underwent CB ablation as index procedure were enrolled. Exclusion criteria were any contraindication for the procedure including the presence of intracavitary thrombus and uncontrolled heart failure and contraindications to general anesthesia.

RESULTS

Acute pericarditis following CB ablation occurred in 18 patients (4%) of our study population. Pericardial effusion occurred in 14 patients (78%) and was mild/moderate. The total number of cryoapplications and the total freeze duration were significantly higher in patients with pericarditis compared with those without (respectively, p = 0.0006 and p = 0.01). Specifically, the number of applications and freeze duration in right inferior pulmonary vein were found significantly higher in patients with pericarditis (p = 0.007). The recurrence rate did not significantly differ between the two study groups (respectively, 16.7 vs 18.1%; p = 0.9).

CONCLUSIONS

The incidence of acute pericarditis following CB ablation in our study population accounted for 4% and was associated with both total freezing time and number of cryoapplications. The clinical course was favorable in all these patients and the occurrence of acute pericarditis did not affect the outcome during the follow-up period.

Для цитирования: Mugnai Giacomo, de Asmundis Carlo, Iacopino Saverio, Ströker Erwin, Longobardi Massimo, Negro Maria Claudia, De Regibus Valentina, Coutino-Moreno Hugo Enrique, Takarada Ken, Choudhury Rajin, de Torres Juan Pablo Abugattas, Storti Cesare, Brugada Pedro, Chierchia Gian-Battista. Acute pericarditis following second-generation cryoballoon ablation for atrial fibrillation.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2018. DOI: 10.1007/s10840-018-0329-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.