Reliability of an automatic ultrasound system for detecting postpartum urinary retention after vaginal birth.
Аннотация
BACKGROUND
Urinary retention is a common postpartum condition which may lead to short- and long-term complications.Objective. To assess the reliability of a commercially available automatic bladder volume scanner in the puerperium, and to specifically identify women with a post-void residual volume of 400 ml or more.
METHODS
A prospective comparison of ultrasound estimated bladder volume and urine volume measurement after catheterisation in 100 women at risk for postpartum urinary retention (PUR) after vaginal delivery.
RESULTS
The mean difference between the ultrasound estimates and the catheter volume measurements was 26 ml, with the corresponding 95% confidence interval (2.6, 49.4 ml). The sensitivity and specificity for the scanner, using the clinically desired value of a 400-ml threshold, are 0.76 and 0.96,respectively. With a perfect classification for sensitivity at a 300-ml threshold, specificity may drop to about 50%. Even using this cut-off point, the number of women catheterised would be reduced by half compared to a procedure of catheterising all women at risk for postpartum urine retention diagnosed by clinical judgment alone.
CONCLUSION
The ultrasound scanner is a reliable screening instrument for detecting PUR after vaginal birth.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.