DOI 10.1109/tbme.2025.3550179

Медицинская библиотека · IEEE transactions on bio-medical engineering · 2025

Design of Magnetic Actuated Capsule Robot for Liquid Sampling.

Zhang Shuo, Song Shaohui, Yu Xinkai, Song Shuang, Zhang Lihai

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

This study aims to introduce a biopsy capsule robot based on the negative pressure suction principle to achieve liquid sampling in the digestive tract.

METHODS

The proposed capsule robot is designed with a magnetic spring configuration. By controlling the direction of an external magnetic field, the suction port can be aligned with the target sampling area. The sampling operation can then be achieved by increasing the external field to start the magnetic spring and a negative pressure can be generated to achieve the liquid sampling. Moreover, a locking mechanism is designed to prevent the magnetic spring from retracting, ensuring that the collected liquid is not squeezed out.

RESULTS

The capsule robot prototype has dimensions of $\phi$ 16.3 mm × 24.4 mm. Both phantom and in-vitro experiments have been carried out. Results showed that it can sample liquids with viscosities ranging from 0.7 mPa$\cdot$s to 200 mPa$\cdot$s and absorb up to 0.24 ml liquid. Additionally, the sealing of the capsule also meets clinical requirements.

CONCLUSION

The experimental results indicate that the designed capsule robot can satisfy the clinical requirements for liquid sampling within the digestive tract.

SIGNIFICANCE: This study has designed and developed a micro capsule robot in the digestive tract, which can achieve safe and efficient liquid sampling operations. The proposed robot can benefit the clinical diagnosis of digestive diseases, especially in the small intestine.

Для цитирования: Zhang Shuo, Song Shaohui, Yu Xinkai, Song Shuang, Zhang Lihai. Design of Magnetic Actuated Capsule Robot for Liquid Sampling.. IEEE transactions on bio-medical engineering. 2025. DOI: 10.1109/tbme.2025.3550179.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.