DOI 10.1016/j.athoracsur.2017.10.001
Mitral Regurgitation Exacerbation Due to Sutureless Aortic Valve Replacement.
Аннотация
Sutureless aortic valves are increasingly used for aortic valve replacement (AVR) with excellent outcomes. Implantation requires device expansion in a subannular position within the native aortic root. We report a patient with severe aortic stenosis who received a Perceval sutureless AVR (LivaNova, London, United Kingdom), resulting in a competent prosthesis with an absence of paravalvular leak and aortic regurgitation. However, this implantation resulted in the exacerbation of mitral regurgitation. The sutureless valve required explantation, and a traditional sutured valve was subsequently implanted, which returned the mitral regurgitation to baseline. We discuss a potential etiology of this sutureless AVR-specific complication.