PURPOSE: To compare laser in situ keratomileusis (LASIK) results obtained with the femtosecond laser (IntraLase Corp.) to those obtained using 2 popular mechanical microkeratomes.

SETTING

Private practice, Greensboro, North Carolina, USA.

METHODS

This retrospective analysis compared LASIK outcomes with the femtosecond laser to those with the Carriazo-Barraquer (CB) microkeratome (Moria, Inc.) and the Hansatome microkeratome (Bausch & Lomb, Inc.). The 3 groups were matched for enrollment criteria and were operated on under similar conditions by the same surgeon.

RESULTS

There were 106 eyes in the IntraLase group, 126 eyes in the CB group, and 143 eyes in the Hansatome group. One day postoperatively, the uncorrected visual acuity (UCVA) results in the 3 groups were similar; at 3 months, the UCVA and the best spectacle-corrected visual acuity results were not significantly different. A manifest spheroequivalent of +/-0.50 diopter (D) was achieved in 91% of eyes in the IntraLase group, 73% of eyes in the CB group, and 74% of eyes in the Hansatome group (P<.01). IntraLase flaps were significantly thinner (P<.01) and varied less in thickness (P<.01) than flaps created with the other devices. The mean flap thickness was 114 microm +/- 14 (SD) with the IntraLase programmed for a 130 microm depth, 153 +/- 26 microm with the CB using a 130 microm plate, and 156 +/- 29 microm with the Hansatome using a 180 microm plate. Loose epithelium was encountered in 9.6% of eyes in the CB group and 7.7% of eyes in the Hansatome group but in no eye in the IntraLase group (P =.001). Surgically induced astigmatism in sphere corrections was significantly less with the IntraLase than with the other devices (P<.01).

CONCLUSIONS

The IntraLase demonstrated more predictable flap thickness, better astigmatic neutrality, and decreased epithelial injury than 2 popular mechanical microkeratomes.

Для цитирования: Kezirian Guy M, Stonecipher Karl G. Comparison of the IntraLase femtosecond laser and mechanical keratomes for laser in situ keratomileusis.. Journal of cataract and refractive surgery. 2004. DOI: 10.1016/j.jcrs.2003.10.026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.