DOI 10.2337/dc20-2856

Медицинская библиотека · Diabetes care · 2021

Accuracy of Dexcom G6 Continuous Glucose Monitoring in Non-Critically Ill Hospitalized Patients With Diabetes.

Davis Georgia M, Spanakis Elias K, Migdal Alexandra L, Singh Lakshmi G, Albury Bonnie, Urrutia Maria Agustina, Zamudio-Coronado K Walkiria, Scott William H, Doerfler Rebecca, Lizama Sergio, Satyarengga Medha, Munir Kashif, Galindo Rodolfo J, Vellanki Priyathama, Cardona Saumeth, Pasquel Francisco J, Peng Limin, Umpierrez Guillermo E

СтатьяПоддержка NIH: внешняя программаПоддержка правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

Advances in continuous glucose monitoring (CGM) have transformed ambulatory diabetes management. Until recently, inpatient use of CGM has remained investigational, with limited data on its accuracy in the hospital setting.

RESEARCH DESIGN AND

METHODS

To analyze the accuracy of Dexcom G6, we compared retrospective matched-pair CGM and capillary point-of-care (POC) glucose data from three inpatient CGM studies (two interventional and one observational) in general medicine and surgery patients with diabetes treated with insulin. Analysis of accuracy metrics included mean absolute relative difference (MARD), median absolute relative difference (ARD), and proportion of CGM values within 15, 20, and 30% or 15, 20, and 30 mg/dL of POC reference values for blood glucose >100 mg/dL or ≤100 mg/dL, respectively (% 15/15, % 20/20, % 30/30). Clinical reliability was assessed with Clarke error grid (CEG) analyses.

RESULTS

A total of 218 patients were included (96% with type 2 diabetes) with a mean age of 60.6 ± 12 years. The overall MARD (n = 4,067 matched glucose pairs) was 12.8%, and median ARD was 10.1% (interquartile range 4.6, 17.6]. The proportions of readings meeting % 15/15, % 20/20, and % 30/30 criteria were 68.7, 81.7, and 93.8%, respectively. CEG analysis showed 98.7% of all values in zones A and B. MARD and median ARD were higher in the case of hypoglycemia (<70 mg/dL) and severe anemia (hemoglobin <7 g/dL).

CONCLUSIONS

Our results indicate that CGM technology is a reliable tool for hospital use and may help improve glucose monitoring in non-critically ill hospitalized patients with diabetes.

Для цитирования: Davis Georgia M, Spanakis Elias K, Migdal Alexandra L, Singh Lakshmi G, Albury Bonnie, Urrutia Maria Agustina, Zamudio-Coronado K Walkiria, Scott William H, Doerfler Rebecca, Lizama Sergio, Satyarengga Medha, Munir Kashif, Galindo Rodolfo J, Vellanki Priyathama, Cardona Saumeth, Pasquel Francisco J, Peng Limin, Umpierrez Guillermo E. Accuracy of Dexcom G6 Continuous Glucose Monitoring in Non-Critically Ill Hospitalized Patients With Diabetes.. Diabetes care. 2021. DOI: 10.2337/dc20-2856.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.