Impact of postoperative patient-prosthesis mismatch as a risk factor for early structural valve deterioration after aortic valve replacement with Trifecta bioprosthesis.
Аннотация
BACKGROUND
Several studies have reported high rates of structural valve deterioration (SVD) in the Trifecta valves. Herein, we analyzed the midterm results of the Trifecta valve and risk factors for early SVD.
METHODS
We retrospectively reviewed the records of 110 patients who had undergone Trifecta implantation between January 2012 and December 2017.
RESULTS
We encountered seven cases of Trifecta valve failure. We performed a redo aortic valve replacement in five patients and a transcatheter aortic valve replacement in two patients. The SVD rate was 4.8% at 5 years and 6.6% at 7 years. The mean pressure gradient and peak velocity on the first postoperative echocardiogram in patients with SVD were higher than those in patients without SVD. The SVD rates with and without patient-prosthesis mismatch (PPM) were 2.8% and 12.6% at 5 years and 2.8% and 20.0% at 7 years. PPM is a risk factor for SVD. Noncoronary cusp tears were observed in all patients who had undergone redo surgery.
CONCLUSIONS
The most common cause of SVD was noncoronary cusp tear. Patients with PPM are at high risk of developing SVD.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.