MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.30301

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2022

Impact of left ventricular outflow tract calcium on valve geometry in self-expanding transcatheter aortic-valve replacement.

Medranda Giorgio A, Rogers Toby, Fuery Michael, Case Brian C, Collins Erin C, Ali Syed W, Zhang Cheng, Wang John C, Weissman Gaby, Cohen Jeffrey E, Shults Christian C, Satler Lowell F, Ben-Dor Itsik, Waksman Ron

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

Evaluate transcatheter heart valve (THV) geometry according to left ventricular outflow tract (LVOT) calcium degree and its impact on hemodynamics and outcomes in patients undergoing transcatheter aortic valve replacement (TAVR) with a contemporary self-expanding THV.

BACKGROUND

LVOT calcium remains challenging for contemporary THVs. LVOT calcium location and degree may affect THV deployment and impact flow patterns and shear stress, accelerating THV degeneration.

METHODS

EPROMPT (CoreValve Evolut Pro Prospective Registry; NCT03423459) is a prospective, investigator-initiated, multicenter registry of patients undergoing TAVR using CoreValve Evolut PRO/PRO + THVs. A total of 107 patients were enrolled in EPROMPT's computed tomography (CT) cohort between January 2018 and October 2021. These patients underwent follow-up CT scan 30 days post-TAVR. We analyzed THV geometry and its interaction with the aortic root following deployment using 30-day post-TAVR CT in patients with none/mild versus moderate/severe LVOT calcium.

RESULTS

Thirty-day THV inflows were less eccentric in the short axis in patients with none/mild versus moderate/severe LVOT calcium (1.16 ± 0.09 vs. 1.21 ± 0.12; p = 0.007). THV became more circular and was similar between both cohorts at the THV waist (1.08 ± 0.06 vs. 1.09 ± 0.11; p = 0.551), leaflet tips (1.03 ± 0.04 vs. 1.05 ± 0.09; p = 0.299), and THV outflow (1.04 ± 2.2 vs. 1.03 ± 2.7; p = 0.143). Thirty-day > mild paravalvular leak was low in both cohorts (1.5% vs. 2.4%; p = 0.724); mean gradients were similar (7.7 ± 3.6 vs. 7.7 ± 3.4 mmHg; p = 0.955).

CONCLUSIONS

Despite inflow deformities observed in patients with moderate/severe LVOT calcium, Evolut PRO/PRO + conforms to elliptical aortic annuli, maintaining circularity and proper function at the leaflets and outflow, even in patients with moderate/severe LVOT calcium.

Для цитирования: Medranda Giorgio A, Rogers Toby, Fuery Michael, Case Brian C, Collins Erin C, Ali Syed W, Zhang Cheng, Wang John C, Weissman Gaby, Cohen Jeffrey E, Shults Christian C, Satler Lowell F, Ben-Dor Itsik, Waksman Ron. Impact of left ventricular outflow tract calcium on valve geometry in self-expanding transcatheter aortic-valve replacement.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2022. DOI: 10.1002/ccd.30301.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.